Flow Antibody Titration Calculator
Plan the dilution series for a new antibody, in µg or µL per test, with the pipetting steps. After the run, paste the statistics table to get the stain index for each tube and the amount to use.
1 · ANTIBODY SECTION 07 — CYTOMETRY · 7.2
Datasheets give either µg per test (enter the stock in mg/mL) or µL per test (the stock can stay empty). A “test” is usually 106 cells in 100 µL; check yours.
2 · SERIES TUBES TO SET UP
3 · STAIN INDEX PASTE THE RESULTS
Amounts are read from sample names (“1:200”, “0.25 µg”, “2 µL”, “4x”) or, failing that, taken in order from the series above. Type over any of them.
| SAMPLE | AMOUNT | POS MFI | NEG MFI | NEG SD | INDEX |
|---|
4 · MY ANTIBODIES SAVED IN THIS BROWSER
| ANTIBODY | LOT | STOCK | TITER | Actions |
|---|
This browser is not keeping data (private window or storage blocked). Export the list to keep it.
How to use
- Enter the ANTIBODY: the stock concentration from the vial and the amount the datasheet recommends per test.
- Set the SERIES: staining volume, how much cell suspension you add, where to start and the step. 4× the recommendation, 2-fold, 8 tubes is a common start.
- Choose SERIAL TRANSFER (one mix, carried tube to tube) or TUBE BY TUBE. Print the plan and tick the tubes off.
- Stain, acquire, and gate the positive and negative populations in each tube. Export the median of both and the robust SD (rSD) of the negative.
- Paste that table under STAIN INDEX. Check the columns and amounts the page picked.
- Save the amount to MY ANTIBODIES. The Staining Cocktail page uses it for every panel after.
The formula
| MFI | median fluorescence of the positive and negative population |
| rSDneg | robust SD of the negative population (SD also works) |
| Separation | (MFIpos − MFIneg) ÷ ((P84neg − MFIneg) ÷ 0.995) |
| Mix | each tube's antibody mix is V ÷ (V − Vc) × the final concentration |
The page marks the highest index and the lowest amount within 90% of it.
Example. 0.2 mg/mL PE antibody, 0.25 µg/test recommended, 100 µL with 50 µL cells, 4× start, 2-fold, 8 tubes, 10% spare: every tube needs 55 µL of mix. Tube 1 is 11 µL antibody + 99 µL buffer; 55 µL goes from tube to tube into 55 µL buffer. Amounts run 1 → 0.0078 µg/test (1:20 → 1:2560).
Good to know
- Free to use. Nothing you paste leaves your browser.
- A titer holds for about the same number of cells in the same volume, at the same temperature and time.
- Titrate viability dyes and tandem dyes the same way; a new lot of a tandem is worth re-checking.
- A calculator, not a protocol: check your lab's method.
Where the numbers come from
| What | Used here | Source |
|---|---|---|
| Stain index | D ÷ W, W = 2 SD of the negative population | Maecker & Trotter, Nature Methods app note 2008, after Maecker et al., Cytometry A 62:169–173 (2004) |
| Robust SD | rSD of the negative as W ÷ 2 | De Rosa, antibody titrations (Fred Hutch / UW CFAR) |
| Separation index | 84th percentile of the negative | FCS Express knowledge base; Telford et al., Cytometry A 2009 (doi) |
| Series | 2-fold, 8 tubes, 4× start | BD Biosciences (4× the recommendation, 2-fold); De Rosa (8–12 dilutions, 106 cells, 50 µL cells + 50 µL antibody) |
| 90% of best | lowest amount within 90% of the top index | FCS Express optimum at 90% of the maximum |
LAST UPDATED · HOW VALUES ARE CHECKED · ABOUT LABMATE.TOOLS
Frequently asked
What is the stain index?
The difference between the positive and negative median fluorescence divided by twice the spread (SD, best the robust SD) of the negative population, as defined by Maecker et al., Cytometry A 2004. It rises with separation and falls when the negative population spreads.
Which concentration should I pick?
The page marks the amount with the highest index and the lowest amount within 90% of it. The lower one saves antibody and usually keeps background down; the best gives the widest separation.
Where do I start the series?
A common start is 2–4× the vendor's recommended amount in 2-fold steps over 8 tubes, which shows the saturation plateau and where the signal drops. Without a recommendation, 1–2 µg per 100 µL is a usual start.
My table has no SD column. Can I still use it?
Yes. With an 84th percentile of the negative population the page uses the separation index; with only medians it falls back to positive ÷ negative, which ignores spread. Export the robust SD from FlowJo to get the stain index.
Does the titer hold for more cells or another volume?
It holds for about the same cell number in the same staining volume. Staining many more cells, or in a different volume, needs the titration repeated or the amount scaled.
Is my data uploaded?
No. The table is read and worked out in your browser. Your antibody list is kept in this browser only; export it as a file to keep a copy.