Lay your markers onto your own cytometer and see, before you stain, where every colour will spill.
| # | MARKER | EXPRESSION | FLUOR | DYE 1–5 | DETECTOR | RECEIVES (MAX) | CHECK | Remove |
|---|
Each preset is a published default or core-facility configuration, linked to the sheet it was taken from. Any filter can be edited, and you can add your own instrument.
| Cytometer | Configuration | Type | Lasers | Detectors | Source |
|---|---|---|---|---|---|
| BD FACSCanto II | 4-2-2 (3 lasers, 8 colors) | Conventional | 405 · 488 · 633 nm | 8 | config sheet |
| BD LSRFortessa X-20 | 5 lasers UV-V-B-YG-R (example core configuration) | Conventional | 355 · 405 · 488 · 561 · 640 nm | 17 | config sheet |
| BD FACSCelesta | 3 lasers V-B-YG, 12 colors (example core configuration) | Conventional | 405 · 488 · 561 nm | 12 | config sheet |
| Beckman Coulter CytoFLEX S | V4-B2-Y4-R3 (13 colors) (check) | Conventional | 405 · 488 · 561 · 638 nm | 13 | config sheet |
| Thermo Fisher Attune NxT | 4 lasers V-B-Y-R, 14 colors (default filters) | Conventional | 405 · 488 · 561 · 637 nm | 14 | config sheet |
| Miltenyi MACSQuant Analyzer 10 | 3 lasers V-B-R, 8 colors | Conventional | 405 · 488 · 640 nm | 8 | config sheet |
| Cytek Aurora | 5 lasers 16UV-16V-14B-10YG-8R (spectral) | Spectral | 355 · 405 · 488 · 561 · 640 nm | 64 | config sheet |
| BD FACSLyric | 3 lasers V·B·R, 12 colors (RUO filter guide) | Conventional | 405 · 488 · 640 nm | 12 | config sheet |
| BD FACSymphony A3 | 5 lasers UV·V·B·YG·R, 28 colors (core example) | Conventional | 355 · 405 · 488 · 561 · 637 nm | 28 | config sheet |
| BD Accuri C6 Plus | 2 lasers B·R, 4 colors (standard) | Conventional | 488 · 640 nm | 4 | config sheet |
| BD FACSAria III | 5 lasers UV·V·B·YG·R sorter (core example) | Conventional | 355 · 405 · 488 · 561 · 640 nm | 15 | config sheet |
| Beckman Coulter CytoFLEX LX | 6 lasers NUV·V·B·Y·R·IR, 21 colors (core example) | Conventional | 375 · 405 · 488 · 561 · 638 · 808 nm | 21 | config sheet |
| Cytek Northern Lights | 3 lasers V16-B14-R8 (spectral) (check) | Spectral | 405 · 488 · 640 nm | 38 | config sheet |
Presets marked (check) follow the maker's listing, but the filter-to-channel mapping still needs checking against a real instrument.
199 dyes from every major supplier, side by side — including Brilliant Violet, Brilliant UltraViolet, StarBright, cFluor, Alexa Fluor, NovaFluor, tandem dyes, viability and proliferation dyes, and fluorescent proteins.
| Excited by | Dyes | Examples |
|---|---|---|
| UV 355 nm | 24 | Spark UV 387, cFluor UV395, BUV395, CellTrace Blue, Sirius, cFluor UV440, … |
| Violet 405 nm | 59 | cFluor V420, Alexa Fluor 405, BV421, Super Bright 436, StarBright Violet 440, cFluor V435, … |
| Blue 488 nm | 39 | EGFP, mEmerald, NovaFluor Blue 510, BB515, cFluor B515, Alexa Fluor 488, … |
| Yellow-green 561 nm | 40 | Alexa Fluor 555, NovaFluor Yellow 570, Alexa Fluor 546, cFluor BYG575, PE, CellTrace Yellow, … |
| Red 640 nm | 36 | Alexa Fluor 633, Alexa Fluor 635, cFluor R659, NovaFluor Red 660, APC, CellTrace Far Red, … |
| 808 nm | 1 | LIVE/DEAD Fixable Near-IR (876), … |
Example. On the Thermo Fisher Attune NxT preset (4 lasers, 14 colours), the sample 9-colour T-cell panel puts CD25 on APC (RL1, 670/14) and CD127 on Alexa Fluor 700 (RL2, 720/30). APC spills an estimated 49% of its primary signal into RL2, so the CD127 row is flagged CHECK. SUGGEST then ranks other dyes for that row by detector clash, spill sent and received, and brightness against expression.
LAST UPDATED · HOW VALUES ARE CHECKED · ABOUT LABMATE.TOOLS
No. Spillover here is a theoretical estimate from spectra and your filters. Real values also depend on detector voltages and laser power, so compensation always comes from single-stain controls on your instrument. Spreading error is not modelled.
Facilities often swap filters. Check the labels on your instrument and use EDIT FILTERS, or add your own instrument.
No. Dyes from every supplier are listed side by side, with peak values from each maker's published tables. Nothing is sponsored.
In your browser. Clearing site data or switching browsers loses them, so export a copy if they matter. You can also share a panel by link.
BD FACSCanto II, LSRFortessa X-20, FACSCelesta, FACSLyric, FACSymphony A3, Accuri C6 Plus and FACSAria III; Beckman Coulter CytoFLEX S and CytoFLEX LX; Thermo Fisher Attune NxT; Miltenyi MACSQuant Analyzer 10; and the spectral Cytek Aurora and Northern Lights. You can also add your own instrument and filters.
Yes. For spectral instruments it compares each dye's signature across all detectors and shows how similar two dyes are, alongside the estimated spillover.
Measured spectra from FPbase (CC BY-SA 4.0). Dyes not on FPbase are modelled from their published excitation and emission peaks. About 200 dyes from every major supplier are listed.
Built-in values are the vendor's own relative rating, currently the BioLegend Fluorophore Brightness Index (1 = dim, 5 = brightest) for the dyes it lists. Each vendor rates its own conjugates on its own scale, so ratings are not absolute and not comparable one-to-one between vendors; real brightness also changes with instrument, filters, clone and F/P ratio. Hover the DYE 1–5 cell to see the source, or pick a number to use your own rating. Dyes without a published rating show — and SUGGEST says so.
It ranks up to four dyes for an open row, or one marked CHECK, by detector clash, spill sent and received, brightness against the marker's expression, and how well the instrument excites and detects them. Brightness counts only for dyes with a known value (a few built-in dyes, or your own rating); each suggestion says which it used. The four are alternatives for one row: two that need the same detector are marked as either-or.