Microscopy Nyquist Calculator

Pick your objective, camera or confocal by brand and model — or type the numbers yourself. Get the resolution, the pixel size and z-step that meet Nyquist, and whether your current settings under- or oversample.

1 · MICROSCOPE SECTION 08 — IMAGING · 8.2

MODE
fTLSTAND · TUBE LENS

Objectives are designed for their own maker's tube lens. Leave this on same maker unless the objective sits on another maker's stand.

2 · OBJECTIVE PICK OR TYPE

MMAGNIFICATION
×
NANUMERICAL APERTURE
nIMMERSION
fdDESIGNED FOR
mm
FNFIELD NUMBER
mm

Engraved on the barrel: 60x/1.42 Oil is magnification / NA / immersion. Field number is on the eyepiece or the objective's spec sheet (often 22–26.5 mm); leave it empty if unsure.

3 · DETECTOR CAMERA

pCAMERA PIXEL
µm
NSENSOR PIXELS
×
bBINNING
McCAMERA ADAPTER
×

Camera adapter = everything between the stand's image and the camera: C-mount adapter, magnification changer, and on a spinning disk the relay after the disk. 1× if none.

4 · CHANNELS DYES & WAVELENGTHS

DYE / PROTEINEXCITATION nmEMISSION nmRemove

Picking a dye fills its peak wavelengths. Type the laser line or filter centre you really use if it differs.

5 · Z STACK OPTIONAL

ΔzYOUR Z STEP
µm
DSTACK DEPTH
µm

6 · RESULT SAMPLING CHECK

SAMPLING

CHANNELEX / EM nmRESOLUTION XYRESOLUTION ZNYQUIST XYNYQUIST Z

Your objective, camera or confocal isn't listed?

You can already use it: type its numbers above and the result is exact. Send it to us and we'll check it against the maker's spec sheet and add it to the shared library, so the next person can pick it from the list.

SUBMIT TO THE LIBRARY

7 · MY HARDWARE SAVED IN THIS BROWSER

TYPENAMEVALUESActions
ABOUT THIS TOOL

How to use

  1. Choose the MODE: widefield, point-scanning confocal, spinning disk or two-photon.
  2. Pick the OBJECTIVE by brand and model, or type magnification, NA and immersion from the barrel.
  3. Pick the CAMERA (widefield, spinning disk) or the CONFOCAL with your frame size and zoom. Or tick I already know the pixel size.
  4. Add your CHANNELS. The strictest one (shortest wavelength) sets the target.
  5. Read the RESULT: pixel size against Nyquist, the adapter, binning or zoom that reaches it, and the z-step. Print the card for the microscope.

The formulas

Δx = λ ÷ (4k · n sin α) · Δz = λ ÷ (2k · n (1 − cos α))
αhalf-aperture angle, sin α = NA ÷ n
Widefieldk = 1, λ = emission
Confocal, diskk = 2, λ = excitation
Two-photonk = 2, λ = the infrared excitation
Pixelcamera: pixel × binning ÷ (M × fTL/fd × Mc); confocal: (F ÷ √2) ÷ (M × zoom × N)

Resolution shown is the Rayleigh criterion (0.61 λ ÷ NA widefield; the pinhole-dependent confocal form).

Example. 60×/1.42 oil, ORCA-Fusion (6.5 µm), 1× adapter, EGFP (507 nm emission), widefield: pixel 108 nm, Nyquist 89 nm — undersampled 1.21×. A 1.5× adapter gives 72 nm, at Nyquist. Same objective on a confocal at 1024 × 1024 with a 20 mm scan field needs zoom 5.4 for the 43 nm target at 488 nm.

Good to know

  1. Free to use. Nothing leaves your browser.
  2. Nyquist is the coarsest sampling that keeps all the resolution; finer costs photons, time and file size.
  3. Real pixel sizes differ from nominal by a few percent. Calibrate with a stage micrometer before you measure.
  4. Hardware values come from the makers' public spec sheets; check yours.
REFERENCE · NYQUIST SAMPLING

Where the numbers come from

WhatUsed hereSource
Nyquist distancesΔx = λ/(4kn sin α), Δz = λ/(2kn(1 − cos α)); λem widefield, λex confocalSVI, Nyquist rate (eqs. 1, 4, 5); SVI FAQ
Tolerated undersamplingabout 1.3× widefield, 1.6× confocal at 1 AUSVI FAQ; UQ IMB deconvolution guide
ResolutionRayleigh, widefield 0.61 λem/NA, 2nλem/NA²; confocal Clat(AU), Cax(AU) with λexNikon MicroscopyU resolution calculator (eqs. 8, 10, 17–20)
Check value63×/1.4 oil, 488 nm confocal → 43 nm / 130 nmUQ IMB deconvolution guide, table 3
Immersion indexwater 1.33, silicone 1.40Evident objectives

Built-in hardware

31 objectives, 15 cameras and 15 confocal and spinning-disk systems, each linked to the page its numbers were read from. Data: CC BY 4.0, compiled 2026-11-24. Not sponsored by any maker.

Objectives

ObjectiveMag. / NAImmersionWD mmTube lensSource
Nikon Plan Apo λD 4x4× / 0.20air20200 mmspec
Nikon Plan Apo λD 10x10× / 0.45air4200 mmspec 2
Nikon Plan Apo λD 20x20× / 0.80air0.8200 mmspec
Nikon Plan Apo λD 40x40× / 0.95air0.21200 mmspec
Nikon Plan Apo λD 60x Oil60× / 1.42oil0.15200 mmspec
Nikon Plan Apo λD 100x Oil100× / 1.45oil0.13200 mmspec
Nikon Plan Apo λS 40xC Sil40× / 1.25silicone0.3200 mmspec
Nikon Apo LWD λS 40xC WI40× / 1.15water0.6200 mmspec 2
Nikon SR Apo TIRF 100x Oil100× / 1.49oil0.12200 mmspec
Evident UPLXAPO4X4× / 0.16air13180 mmspec
Evident UPLXAPO10X10× / 0.40air3.1180 mmspec
Evident UPLXAPO20X20× / 0.80air0.6180 mmspec
Evident UPLXAPO40X40× / 0.95air0.18180 mmspec
Evident UPLXAPO40XO40× / 1.40oil0.13180 mmspec
Evident UPLXAPO60XO60× / 1.42oil0.15180 mmspec
Evident UPLXAPO100XO100× / 1.45oil0.13180 mmspec
Zeiss Plan-Apochromat 20x/0.820× / 0.80air0.55165 mmspec
Zeiss Plan-Apochromat 63x/1.4 Oil DIC63× / 1.40oil165 mmspec
Zeiss Plan-Apochromat 100x/1.40 Oil DIC100× / 1.40oil165 mmspec
Zeiss α Plan-Apochromat 100x/1.46 Oil DIC100× / 1.46oil165 mmspec
Zeiss C-Apochromat 10x/0.45 W10× / 0.45water1.8165 mmspec
Zeiss C-Apochromat 40x/1.2 W Corr40× / 1.20water165 mmspec
Zeiss C-Apochromat 63x/1.20 W Corr63× / 1.20water165 mmspec
Zeiss LD C-Apochromat 40x/1.1 W Corr40× / 1.10water0.62165 mmspec
Leica HC PL APO 10x/0.40 CS210× / 0.40air2.56200 mmspec
Leica HC PL APO 20x/0.75 CS220× / 0.75air0.62200 mmspec
Leica HC PL APO 40x/1.30 OIL CS240× / 1.30oil0.24200 mmspec
Leica HC PL APO 63x/1.40 OIL CS263× / 1.40oil0.14200 mmspec
Leica HC PL APO 63x/1.30 GLYC CORR CS263× / 1.30glycerol0.3200 mmspec
Leica HC PL APO 63x/1.20 W CORR CS263× / 1.20water0.3200 mmspec
Leica HC PL APO 100x/1.40 OIL STED WHITE100× / 1.40oil0.13200 mmspec

Cameras

CameraSensorPixel µmPixelsSource
Hamamatsu ORCA-Fusion BT (C15440-20UP)BSI sCMOS6.52304 × 2304spec
Hamamatsu ORCA-Fusion (C14440-20UP)sCMOS6.52304 × 2304spec
Hamamatsu ORCA-Flash4.0 V3 (C13440-20CU)sCMOS6.52048 × 2048spec
Hamamatsu ORCA-Quest (C15550-20UP)qCMOS4.64096 × 2304spec
Hamamatsu ORCA-Fire (C16240-20UP)BSI sCMOS4.64432 × 2368spec
Teledyne Photometrics Prime BSI ExpressBSI sCMOS6.52048 × 2048spec
Teledyne Photometrics Prime BSI (check)BSI sCMOS6.52048 × 2048spec
Teledyne Photometrics KinetixBSI sCMOS6.53200 × 3200spec
Teledyne Photometrics Prime 95B (check)BSI sCMOS111200 × 1200spec
Andor Zyla 4.2 / ZL41 Cell 4.2 (check)sCMOS6.52048 × 2048spec
Andor Zyla 5.5 / ZL41 Cell 5.5 (check)sCMOS6.52560 × 2160spec
Andor Sona-6 (4.2B-6) (check)BSI sCMOS6.52048 × 2048spec
Andor Sona-11 (4.2B-11) (check)BSI sCMOS112048 × 2048spec
Andor iXon Life 888 (check)EMCCD131024 × 1024spec
Andor iXon Life 897 (check)EMCCD16512 × 512spec

Confocal and spinning disk

SystemTypeField / pinholeMax frameSource
Zeiss LSM 880 (discontinued)point scan20 mm field diagonal in the intermediate image8192 pxspec
Zeiss LSM 980point scan20 mm field diagonal in the intermediate image8192 pxspec
Leica TCS SP8 (FOV scanner) (discontinued)point scanField diameter 22 mm8192 pxspec
Leica TCS SP8 (resonant 8 kHz) (discontinued)point scanField diameter 13 mm2496 pxspec
Leica STELLARIS 8point scan22 mm (FOV) diagonal in the intermediate image plane8192 pxspec
Leica STELLARIS 8 (resonant)point scan13 mm diagonal2496 pxspec
Nikon A1 HD25 (discontinued)point scanSquare inscribed in a Ø25 mm circle (Ti2-E)4096 pxspec
Nikon A1R HD25 (resonant) (discontinued)point scanSquare inscribed in a Ø25 mm circle (Ti2-E)1024 pxspec
Nikon AX (galvano)point scanField Number 258192 pxspec
Nikon AX R (resonant)point scanField Number 252048 pxspec
Evident FV3000 (galvano) (check)point scanField number 18 (to check)4096 pxspec
Evident FV4000 (galvano) (check)point scanField number 18 (to check)4096 pxspec
Yokogawa CSU-W1 · 50 µm disk (check)spinning disk50 µm pinholes, 500 µm apartspec
Yokogawa CSU-W1 · 25 µm disk (check)spinning disk25 µm pinholes, 500 µm apartspec
Yokogawa CSU-X1 (check)spinning disk50 µm pinholes, 253 µm apartspec

(check) = at least one number derived or commonly cited, still to confirm against the maker. Dye peaks are shared with the Panel Designer list.

LAST UPDATED · HOW VALUES ARE CHECKED · ABOUT LABMATE.TOOLS

QUESTIONS · NYQUIST SAMPLING

Frequently asked

What pixel size do I need for Nyquist sampling?

For widefield, the pixel on the sample should be at most λem ÷ (4 NA); for a confocal, λex ÷ (8 NA). With NA 1.4 and GFP that is about 91 nm widefield and 44 nm confocal. The page works it out per channel and uses the strictest.

My objective or camera is not in the list. Can I still use it?

Yes. Every field can be typed: magnification, NA, immersion, camera pixel size, frame size, zoom, wavelengths. The lists only fill the numbers in. Use SUBMIT TO THE LIBRARY to send us your model so it can be added for everyone.

Is undersampling always bad?

No. You lose resolution you paid for in NA, but you gain field of view, speed and signal per pixel. For deconvolution, SVI suggests staying within about 1.3× (widefield) to 1.6× (confocal at 1 Airy unit) of the Nyquist distance.

Why does the tube lens matter?

Infinity objectives are designed for one tube lens: 200 mm for Nikon and Leica, 180 mm for Evident and 165 mm for Zeiss. On another maker's stand the real magnification is the nominal one × stand tube lens ÷ design tube lens.

Where do the hardware specifications come from?

From each maker's public specification pages, datasheets and core-facility documents, linked per entry in the table below. Entries marked check have a number still to confirm. The list is independent and not sponsored.

Is anything uploaded?

No. Everything is worked out in your browser. Your own objectives and cameras are kept in this browser; export them to keep a copy.